Differential Pressure Testing
Our DP wands have a locking system, plunger or bellows that lock the DP wand into the tube to prevent leakage during testing. The primary intent of this set up is to produce, essentially constant flow through each tube during the testing by achieving a critical pressure ratio significantly greater than twice the absolute pressure down stream of the orifice. The expected DP across the tubes is maintained as per requirement.